Climatic test chamber temp. -70°C to +180°CVötsch 190 Liter
VCS 7018-5
Climatic test chamber temp. -70°C to +180°C
Vötsch 190 Liter
VCS 7018-5
Condition
Used
Location
Borken 

Machine data
- Machine type:
- Climatic test chamber temp. -70°C to +180°C
- Manufacturer:
- Vötsch 190 Liter
- Model:
- VCS 7018-5
- Condition:
- excellent (used)
Price & Location
- Location:
- Einsteinstraße 8a, 46325 Borken, Deutschland

Call
Details about the offer
- Listing ID:
- A222-57222
- Reference number:
- 24557
- Update:
- 03.07.2026
Description
Heraeus Vötsch
Type: VCS 7018-5
Climate testing chamber with humidity, water-cooled
Test chamber dimensions: Width 580 mm x Depth 450 mm x Height 750 mm
Housing dimensions: Width 870 mm x Depth 1280 mm x Height 1775 mm
Specifications for temperature tests:
Temperature range: -70°C to +180°C
Temporal accuracy: ± 0.1 to ± 0.5 K
Spatial accuracy: ± 0.5 to ± 2.0 K
Temperature gradient: 1 to 4 K
Temperature change rate according to IEC 60068-3-5
Heating: 7.5 K/min
Cooling: 7.5 K/min
Specifications for climate tests:
Bjdpfx Aeziiwhsmbodd
Temperature range: +10 to +95°C
Humidity range: 10 to 98 % RH
Temporal temperature deviation: ± 0.1 to ± 0.3 K
Spatial temperature deviation: ± 0.5 to ± 1 K
Temperature gradient: 1 to 2 K
Temporal humidity fluctuation: ± 1 to ± 3 % RH
Dew point range: -3 to +94 °C
Electrical connection:
Rated voltage: 3/N/PE AC 400V ±10% 50Hz
Rated power: 10 kW
Rated current: 21 A
Fuse rating: 32 A slow-blow, on-site
Refrigerant: R-404A / R-23
Test chamber volume: approx. 190 liters
Weight: approx. 460 kg
For your safety as a buyer, please note the following:
The following points are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. If necessary, refilling with legally compliant refrigerant
3. Leakage test with certificate
4. After successful inspection, the chambers are subjected to a documented test run.
Condition: used
Scope of delivery: (See picture)
(Changes and errors in the technical data are subject to change!)
We will be happy to answer any further questions you may have by telephone.
The listing was translated automatically. Translation errors are possible.
Type: VCS 7018-5
Climate testing chamber with humidity, water-cooled
Test chamber dimensions: Width 580 mm x Depth 450 mm x Height 750 mm
Housing dimensions: Width 870 mm x Depth 1280 mm x Height 1775 mm
Specifications for temperature tests:
Temperature range: -70°C to +180°C
Temporal accuracy: ± 0.1 to ± 0.5 K
Spatial accuracy: ± 0.5 to ± 2.0 K
Temperature gradient: 1 to 4 K
Temperature change rate according to IEC 60068-3-5
Heating: 7.5 K/min
Cooling: 7.5 K/min
Specifications for climate tests:
Bjdpfx Aeziiwhsmbodd
Temperature range: +10 to +95°C
Humidity range: 10 to 98 % RH
Temporal temperature deviation: ± 0.1 to ± 0.3 K
Spatial temperature deviation: ± 0.5 to ± 1 K
Temperature gradient: 1 to 2 K
Temporal humidity fluctuation: ± 1 to ± 3 % RH
Dew point range: -3 to +94 °C
Electrical connection:
Rated voltage: 3/N/PE AC 400V ±10% 50Hz
Rated power: 10 kW
Rated current: 21 A
Fuse rating: 32 A slow-blow, on-site
Refrigerant: R-404A / R-23
Test chamber volume: approx. 190 liters
Weight: approx. 460 kg
For your safety as a buyer, please note the following:
The following points are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. If necessary, refilling with legally compliant refrigerant
3. Leakage test with certificate
4. After successful inspection, the chambers are subjected to a documented test run.
Condition: used
Scope of delivery: (See picture)
(Changes and errors in the technical data are subject to change!)
We will be happy to answer any further questions you may have by telephone.
The listing was translated automatically. Translation errors are possible.
Send inquiry
Telephone & Fax
+49 2861 ... show
Your listing has been deleted successfully
An error has occurred